Ionization Chambers AEC Sensor (VacuTec, Germany)
Ionization Chambers for Automatic X-Ray Exposure Control
The use of AEC Ionization Chambers for Automatic X-Ray Exposure Control is mandatory according to Article 17 of the Law of Ukraine «On the protection of humans from the impact of ionizing radiation».
Energy range / tube voltage | (40 ... 150) kV |
Dose rate range | (0,5 ... 1000) µGy/s |
Exposure dose range1) | (1 ... 100) µGy |
Digital resolution (selectable) | 0,025 µGy |
Eposure time range1) | 1 ms ... 10 s |
Sensitivity difference between sensor fields | < 5% |
Attenuation factor1) | < 1,04 |
Aluminium equivalent | < 0,75 mm Al |
Supply voltage (positive and negative) | ±(11,5 ... 16)V DC |
Digital output | Differential signal (RS 422), pulse width 2 µs |
Ramp output (with ramp module) | 0 ... 10 V |
1)
According to DIN 6815 dose measurement is carried out as follows: the air kerma (in Gy) is measured immediately downstream of the combined scattered-ray grid/measuring
chamber using a radiation quality attained at a high-voltage of (70...80) kV and an Al attenuation layer thickness of 25 mm. |
For detailed technical information please require our documents. |
Click for larger schemes
Ionization Chambers AEC Sensor (VacuTec)
Application
The ionization chambers are used as sensor elements for automatic X-ray exposure control units.
They are comprised of air-filled parallel chambers with 1, 3 measuring fields, coupled with an amplifier and a field selector.
The low overall thickness of 6 mm ensures the chamber may be used with all modern X-ray equipment.
Principle of Operation
The current generated in the sensor fields by the X-rays will be amplified, digitized and fed to the Automatic Exposure Controller.
The AEC supplies the cut-off signal to the generator, and thereby ensures an optimal and reproducible image quality, independent of the radiation quality, object density and other factors.
Chamber Electronics
The ionization chambers are supplied with minimized chamber electronics, that amplify, digitize and process the generated current according to the field selection.
The electronics also generate the chamber voltage without the need for an external high-voltage source.
Ramp Modul
In order to assure the backward compatibility with conventional interfaces a separate ramp module converts the pulses into a voltage ramp.
This ramp module should be placed between the interface cable and the AEC interface on the generator site.
Chamber Types
The listed sizes of AEC chambers are supplied as a standard, but other dimensions are available on request.
The minimized chamber electronics are mounted directly on the side of the chamber. If necessary there are versions with electronics separated up to 90 cm distance from the ionization chamber.
Type | Description | Size L2xB2 (mm) |
141 00 18 | 3-field chamber with electronics | 374 x 374 |
142 00 13 | 1-field chamber with electronics | 374 x 374 |
145 00 44 | 3-field chamber with electronics | 450 x 450 |
147 00 05 | 3-field chamber with electronics | 370 x 450 |
151 00 21ì | 3-field chamber with electronics | 450 x 470 |
High-level Chamber
The series AEC-Sensor digital has been optimised for latest flat-panel detectors.
Chambers almost invisible measuring fields are a result of an innovative construction, which yields an extremely homogeneous X-ray absorption of the measuring chamber for all relevant tube-voltages in medical X-ray diagnostics.
Chambers thickness of 6 mm, flexibility in size and the specifications correspond to the conventional ionization chambers.
Explanation of the information and advertising nature of the information posted on the site.
Dear visitors of our site. All information contained on the site is for informational and advertising purposes only. The information provided on the site is not exhaustive.Technical specifications may be changed or supplemented at any time.
This information cannot be used as official document, for example, for the preparation of the technical part of the tender documentation.
For more information, contact our experts. Contact Us >>
|